Precurved non-tunnelled catheters for haemodialysis are comparable in terms of infections and malfunction as compared to tunnelled catheters: A retrospective cohort study
BACKGROUND:: The main limitations of central venous catheters for haemodialysis access are infections and catheter malfunction. Our objective was to assess whether precurved non-tunnelled central venous catheters are comparable to tunnelled central venous catheters in terms of infection and catheter malfunction and to assess whether precurved non-tunnelled catheters are superior to straight catheters. MATERIALS AND METHODS:: In this retrospective, observational cohort study, adult patients in whom a central venous catheter for haemodialysis was inserted between 2012 and 2016 were included. The... Mehr ...
Verfasser: | |
---|---|
Dokumenttyp: | Artikel |
Erscheinungsdatum: | 2019 |
Schlagwörter: | catheter / complications / infections / Haemodialysis / vascular access / Risk Assessment / Central Venous Catheters / Humans / Middle Aged / Risk Factors / Catheter-Related Infections/diagnosis / Male / Treatment Outcome / Equipment Design / Netherlands / Catheterization / Central Venous/adverse effects / Female / Aged / Retrospective Studies / Catheters / Indwelling / Equipment Failure / Renal Dialysis / Nephrology / Surgery / Observational Study / Multicenter Study / Journal Article / Comparative Study |
Sprache: | Englisch |
Permalink: | https://search.fid-benelux.de/Record/base-29202542 |
Datenquelle: | BASE; Originalkatalog |
Powered By: | BASE |
Link(s) : | https://dspace.library.uu.nl/handle/1874/388657 |
BACKGROUND:: The main limitations of central venous catheters for haemodialysis access are infections and catheter malfunction. Our objective was to assess whether precurved non-tunnelled central venous catheters are comparable to tunnelled central venous catheters in terms of infection and catheter malfunction and to assess whether precurved non-tunnelled catheters are superior to straight catheters. MATERIALS AND METHODS:: In this retrospective, observational cohort study, adult patients in whom a central venous catheter for haemodialysis was inserted between 2012 and 2016 were included. The primary endpoint was a combined endpoint consisting of the first occurrence of either an infection or catheter malfunction. The secondary endpoint was a combined endpoint of the removal of the central venous catheter due to either an infection or a catheter malfunction. Using multivariable analysis, cause-specific hazard ratios for endpoints were calculated for tunnelled catheter versus precurved non-tunnelled catheter, tunnelled catheter versus non-tunnelled catheter, and precurved versus straight non-tunnelled catheter. RESULTS:: A total of 1603 patients were included. No difference in reaching the primary endpoint was seen between tunnelled catheters, compared to precurved non-tunnelled catheters (hazard ratio, 0.91; 95% confidence interval, 0.70-1.19, p = 0.48). Tunnelled catheters were removed less often, compared to precurved non-tunnelled catheters (hazard ratio, 0.65; 95% confidence interval, 0.46-0.93; p = 0.02). A trend for less infections and catheter malfunctions was seen in precurved jugular non-tunnelled catheters compared to straight non-tunnelled catheters (hazard ratio, 0.60; 95% confidence interval, 0.24-1.50; p = 0.28) and were removed less often (hazard ratio, 0.41; 95% confidence interval, 0.18-0.93; p = 0.03). CONCLUSION:: Tunnelled central venous catheters and precurved non-tunnelled central venous catheters showed no difference in reaching the combined endpoint of catheter-related infections and ...