Midterm clinical outcome following Edwards SAPIEN or Medtronic Corevalve transcatheter aortic valve implantation (TAVI): Results of the Belgian TAVI registry.

peer reviewed ; OBJECTIVE: To assess midterm (3 years) clinical outcomes of transcatheter aortic valve implantation (TAVI) in Belgium using the Edwards SAPIEN valve or the Medtronic CoreValve transcatheter heart valve (THV). BACKGROUND: Medium and long term follow-up data of both THVs are still relatively scarce, although of great clinical relevance for a relatively new but rapidly expanding treatment modality. Therefore, reporting mid- and long term clinical outcome data, coming from large "real world" national registries, remains contributive. METHODS: Between December 2007 and March 2012, 8... Mehr ...

Verfasser: Collas, Valerie M.
Dubois, Christophe
LEGRAND, Victor
Kefer, Joelle
De Bruyne, Bernard
Dens, Jo
Rodrigus, Inez E.
Herijgers, Paul
Bosmans, Johan M.
Dokumenttyp: journal article
Erscheinungsdatum: 2015
Verlag/Hrsg.: John Wiley & Sons
Schlagwörter: aortic stenosis / registry / transcatheter aortic valve implantation (TAVI) / valvular heart disease / Human health sciences / Cardiovascular & respiratory systems / Sciences de la santé humaine / Systèmes cardiovasculaire & respiratoire
Sprache: Englisch
Permalink: https://search.fid-benelux.de/Record/base-28888716
Datenquelle: BASE; Originalkatalog
Powered By: BASE
Link(s) : https://orbi.uliege.be/handle/2268/193267

peer reviewed ; OBJECTIVE: To assess midterm (3 years) clinical outcomes of transcatheter aortic valve implantation (TAVI) in Belgium using the Edwards SAPIEN valve or the Medtronic CoreValve transcatheter heart valve (THV). BACKGROUND: Medium and long term follow-up data of both THVs are still relatively scarce, although of great clinical relevance for a relatively new but rapidly expanding treatment modality. Therefore, reporting mid- and long term clinical outcome data, coming from large "real world" national registries, remains contributive. METHODS: Between December 2007 and March 2012, 861 "real world" patients who were not candidates for surgical aortic valve replacement as decided by the local heart teams, underwent TAVI at 23 sites. Eleven sites exclusively used SAPIEN THV (n = 460), while 12 exclusively used CoreValve THV (n = 401). Differences in clinical outcomes by valve system were assessed, according to access route and baseline EuroSCORE risk profile (<10%: low, 10-20%: intermediate and >20%: high risk). RESULTS: Overall cumulative survival at 3 years was 51% for SAPIEN vs. 60% for CoreValve (P = 0.021). In transfemorally treated patients, SAPIEN and CoreValve had similar survival at 3 years for each of the baseline EuroSCORE cohorts (low risk: 72% vs. 76%, P = 0.45; intermediate risk: 62% vs. 59%, P = 0.94; high risk: 48% vs. 53%, P = 0.65). CONCLUSION: Cumulative midterm 3 year survival after transfemoral TAVI in "real world" patients refused for surgery with similar baseline EuroSCORE risk profile is not different between SAPIEN or CoreValve.