Quality assurance in CT with the Belgian protocol and the new European acceptability criteria
In a working group of the Belgian Hospital Physicists Association (BHPA), a new protocol has been developed for comprehensive testing of computed tomography scanners. The tests were selected to verify if the scanner is technically adequate, if preprogrammed patient protocols are up-to-date, and if exposure values displayed at the console are sufficiently correct. In addition, they will ensure that the participating medical physics expert (MPE) gets a full understanding of the system to enable him/her to guide optimization processes and allow automated patient dosimetry. Several new test proced... Mehr ...
Verfasser: | |
---|---|
Dokumenttyp: | TEXT |
Erscheinungsdatum: | 2013 |
Verlag/Hrsg.: |
Oxford University Press
|
Schlagwörter: | Section 5: Diagnostic Radiology / General and CT |
Sprache: | Englisch |
Permalink: | https://search.fid-benelux.de/Record/base-28887293 |
Datenquelle: | BASE; Originalkatalog |
Powered By: | BASE |
Link(s) : | http://rpd.oxfordjournals.org/cgi/content/short/153/2/197 |