Prevalence and Presentation of Externalized Conductors and Electrical Abnormalities in Riata Defibrillator Leads After Fluoroscopic Screening ; Report From the Netherlands Heart Rhythm Association Device Advisory Committee

Background— The Riata family of implantable cardioverter-defibrillator (ICD) leads is prone to a specific insulation abrasion characterized by externalization of conductor cables. The objective of this study was to determine the prevalence of externalized conductors and electrical abnormalities in Riata ICD leads by fluoroscopic screening and standard ICD interrogation. Methods and Results— All ICD implantation centers were contacted by the Netherlands Heart Rhythm Association Device Advisory Committee to identify all patients with an active Riata ICD lead and to perform fluoroscopic screening... Mehr ...

Verfasser: Theuns, Dominic A.M.J.
Elvan, Arif
de Voogt, Willem
de Cock, Carel C.
van Erven, Lieselot
Meine, Mathias
Dokumenttyp: Artikel
Erscheinungsdatum: 2012
Reihe/Periodikum: Circulation: Arrhythmia and Electrophysiology ; volume 5, issue 6, page 1059-1063 ; ISSN 1941-3149 1941-3084
Verlag/Hrsg.: Ovid Technologies (Wolters Kluwer Health)
Sprache: Englisch
Permalink: https://search.fid-benelux.de/Record/base-27612476
Datenquelle: BASE; Originalkatalog
Powered By: BASE
Link(s) : http://dx.doi.org/10.1161/circep.112.975755

Background— The Riata family of implantable cardioverter-defibrillator (ICD) leads is prone to a specific insulation abrasion characterized by externalization of conductor cables. The objective of this study was to determine the prevalence of externalized conductors and electrical abnormalities in Riata ICD leads by fluoroscopic screening and standard ICD interrogation. Methods and Results— All ICD implantation centers were contacted by the Netherlands Heart Rhythm Association Device Advisory Committee to identify all patients with an active Riata ICD lead and to perform fluoroscopic screening of the lead. In addition, the electrical integrity of the lead was assessed. As of March 1, 2012, data for 1029 active Riata leads were available; 47% of these were 8-F Riata and 53% were 7-F Riata ST. Externalized conductors were observed in 147 leads (14.3%). Proportion of externalized conductors was higher in 8-F Riata compared with 7-F Riata ST (21.4% vs 8.0%; P <0.001). Median time from implantation to detection of externalized conductors was 65.3 months. The estimated rates of externalized conductors were 6.9% and 36.6% at 5 and 8 years after implantation, respectively. Of the 147 leads with externalized conductors, 10.9% had abnormal electrical parameters vs 3.5% in nonexternalized leads ( P <0.001). Conclusions— The prevalence of externalized conductors in Riata leads is significantly high (14.3%) using fluoroscopic screening. The majority of externalized conductors are not detectable with standard ICD interrogation. Screening with fluoroscopy is reasonable.