Sustained suppression of neointimal proliferation by sirolimus-eluting stents: one-year angiographic and intravascular ultrasound follow-up

BACKGROUND: We have previously reported a virtual absence of neointimal hyperplasia 4 months after implantation of sirolimus-eluting stents. The aim of the present investigation was to determine whether these results are sustained over a period of 1 year. METHODS AND RESULTS: Forty-five patients with de novo coronary disease were successfully treated with the implantation of a single sirolimus-eluting Bx VELOCITY stent in Sao Paulo, Brazil (n=30, 15 fast release [group I, GI] and 15 slow release [GII]) and Rotterdam, The Netherlands (15 slow release, GIII). Angiographic and volumetric intravas... Mehr ...

Verfasser: Sousa, J.E. (Eduardo)
Langenhove, G.J.J. (Glenn) van
Sousa, A.G.M.R. (Amanda)
Falotico, R. (Robert)
Popma, J.J. (Jeffrey)
Jaeger, J. (Judith)
Costa, M.A. (Marco)
Abizaid, A.C. (Alexandre)
Rensing, B.J.W.M. (Benno)
Abizaid, A.S. (Andrea)
Tanajura, L.F. (Luiz)
Kozuma, K. (Ken)
Serruys, P.W.J.C. (Patrick)
Dokumenttyp: Artikel
Erscheinungsdatum: 2001
Schlagwörter: *Stents/adverse effects / Blood Vessel Prosthesis Implantation/adverse effects/*instrumentation/methods / Brazil / Cohort Studies / Coronary Angiography / Coronary Disease/diagnosis/drug therapy/*surgery / Delayed-Action Preparations/administration & dosage / Drug Implants/administration & dosage / Endosonography / Female / Follow-Up Studies / Graft Occlusion / Vascular/etiology/*prevention & control / Humans / Male / Middle aged / Netherlands / Sirolimus/*administration & dosage/adverse effects / Survival Rate / Treatment Outcome / Tunica Intima/drug effects / Vascular Patency/drug effects
Sprache: Englisch
Permalink: https://search.fid-benelux.de/Record/base-27218279
Datenquelle: BASE; Originalkatalog
Powered By: BASE
Link(s) : http://repub.eur.nl/pub/9778

BACKGROUND: We have previously reported a virtual absence of neointimal hyperplasia 4 months after implantation of sirolimus-eluting stents. The aim of the present investigation was to determine whether these results are sustained over a period of 1 year. METHODS AND RESULTS: Forty-five patients with de novo coronary disease were successfully treated with the implantation of a single sirolimus-eluting Bx VELOCITY stent in Sao Paulo, Brazil (n=30, 15 fast release [group I, GI] and 15 slow release [GII]) and Rotterdam, The Netherlands (15 slow release, GIII). Angiographic and volumetric intravascular ultrasound (IVUS) follow-up was obtained at 4 and 12 months (GI and GII) and 6 months (GIII). In-stent minimal lumen diameter and percent diameter stenosis remained essentially unchanged in all groups (at 12 months, GI and GII; at 6 months, GIII). Follow-up in-lesion minimal lumen diameter was 2.28 mm (GIII), 2.32 mm (GI), and 2.48 mm (GII). No patient approached the >/=50% diameter stenosis at 1 year by angiography or IVUS assessment, and no edge restenosis was observed. Neointimal hyperplasia, as detected by IVUS, was virtually absent at 6 months (2+/-5% obstruction volume, GIII) and at 12 months (GI=2+/-5% and GII=2+/-3%). CONCLUSIONS: This study demonstrates a sustained suppression of neointimal proliferation by sirolimus-eluting Bx VELOCITY stents 1 year after implantation.