Guidelines for Quality Control of Equipment Used in Diagnostic Radiology in the Netherlands

The Dutch working group on 'Quality Criteria for Equipment Used in Diagnostic Radiology' has formulated guidelines providing technical criteria for equipment used in conventional diagnostic radiology. These guidelines are applicable to the technical parameters having a major impact on image quality and patient dose and include methods for testing. The following parameters are included: tube voltage, automatic exposure control, film processing, film-screen combination, light tightness and illumination of the dark room, half-value layer and filtration, light field, grid, focal spot size, viewing... Mehr ...

Verfasser: van den Berg, L.
Aarts, J.C.N.M.
Beentjes, L.B.
van Dalen, A.
Elsakkers, P.
Julius, H.W.
Kicken, P.J.H.
van der Meer, F.
Teeuwisse, W.
Thijssen, M.A.O.
Zoetelief, J.
Dokumenttyp: TEXT
Erscheinungsdatum: 1998
Verlag/Hrsg.: Oxford University Press
Schlagwörter: Article
Sprache: Englisch
Permalink: https://search.fid-benelux.de/Record/base-26807576
Datenquelle: BASE; Originalkatalog
Powered By: BASE
Link(s) : http://rpd.oxfordjournals.org/cgi/content/short/80/1-3/95