Patient Dose and Image Quality for Computed Tomography in Several Dutch Hospitals

In several Dutch hospitals a survey of computed tomography (CT) techniques was performed at 14 CT scanners. In this survey the computed tomography dose index (CTDI) was measured free-in-air with a pencil type ionisation chamber. These measurements yield dose values related to a single CT scan of one slice whilst radiation exposure of patients due to a complete CT examination should preferably be expressed in terms of organ and tissue doses or effective dose. Organ and effective doses were derived by application of published organ dose conversion factors. Effective doses varied from 0.7 mSv to... Mehr ...

Verfasser: Geleijns, J.
Broerse, J.J.
Zoetelief, J.
Zweers, D.
van Unnik, J.G.
Dokumenttyp: TEXT
Erscheinungsdatum: 1995
Verlag/Hrsg.: Oxford University Press
Schlagwörter: Article
Sprache: Englisch
Permalink: https://search.fid-benelux.de/Record/base-26633723
Datenquelle: BASE; Originalkatalog
Powered By: BASE
Link(s) : http://rpd.oxfordjournals.org/cgi/content/short/57/1-4/129