A LARGE-SCALE MULTICENTRE STUDY IN BELGIUM OF DOSE AREA PRODUCT VALUES AND EFFECTIVE DOSES IN INTERVENTIONAL CARDIOLOGY USING CONTEMPORARY X-RAY EQUIPMENT

In this paper, a large-scale multicentre patient dose study performed in eight Belgian interventional cardiology departments is presented. Effective dose ( E ) was calculated based on a detailed dose–area product (DAP)-registration during each procedure and by using conversion coefficients generated by the Monte Carlo-based computer program PCXMC. Conversion coefficients were found to be 0.177 mSv Gycm−2 for systems that do not use any additional copper filtration in cineradiography and 0.207 mSv Gycm−2 for systems that use additional copper filtration in cineradiography. Mean E values of 9.6... Mehr ...

Verfasser: Bogaert, E.
Bacher, K.
Thierens, H.
Dokumenttyp: TEXT
Erscheinungsdatum: 2007
Verlag/Hrsg.: Oxford University Press
Schlagwörter: SCIENTIFIC AND TECHNICAL PAPER
Sprache: Englisch
Permalink: https://search.fid-benelux.de/Record/base-26614585
Datenquelle: BASE; Originalkatalog
Powered By: BASE
Link(s) : http://rpd.oxfordjournals.org/cgi/content/short/ncm379v1